- Easy to read Color Graphical TFT LCD display
- Rechargeable lithium ion battery with enhanced power control for prolonged battery life
- Measure or Source: mA, mV and V readings
- 24 VDC Loop power Supply to power transmitters and loops
- 2W simulate and Read/ Power mode for in-situ Loop checking and calibration
- Step/ Ramp functions with Auto/ Man selection
- Universal Serial Bus (USB) communication port for charging, data retrieve and firmware upgrade
- Data Logging to measure long time drift
- Other Features: Max/Min/Average, filter settings, tare facility, adjustable backlight, alarm annunciation (on display and buzzer), automatic Display off.
- Easy to read Color Graphical TFT LCD display
- Rechargeable lithium ion battery with enhanced power control for prolonged battery life
- Measure or Source: mA, mV and V readings
- 24 VDC Loop power Supply to power transmitters and loops
- 2W simulate and Read/ Power mode for in-situ Loop checking and calibration
- Step/ Ramp functions with Auto/ Man selection
- Universal Serial Bus (USB) communication port for charging, data retrieve and firmware upgrade
- Data Logging to measure long time drift
- Other Features: Max/Min/Average, filter settings, tare facility, adjustable backlight, alarm annunciation (on display and buzzer), automatic Display off.
- Direct Temperature output 12 Switched Steps
- < 0.3 °C Accuracy
- Quick Check Switch BOX
- Allows 2, 3 or 4 wire connections
- Compact Size
- Compact, handheld, User friendly menu
- Easy to read Color Graphical TFT LCD display
- Rechargeable lithium Ion battery with enhanced power control for prolonged battery life
- Measure: mA/ V/ mV/ mA (24V)/ Switch-test / RTD/ TC/ Frequency/ Pulse
- Source: mA/ V/ mV/ mA(2W)/ Resistance/ RTD/ TC/ Frequency/ Pulse
- 24 VDC Loop power Supply to power transmitters and loops
- Step/Ramp functions with Auto/Man selection
- Universal Serial Bus (USB) communication port for charging, data retrieve and firmware upgrade
- Data Logging to measure long time drift
- Other Features: Max/Min/Average, filter settings, tare facility, adjustable backlight, alarm annunciation (on display and buzzer), glance screen mode
- Continuity Test
- Pulsed RTD transmitter compatible
- HART loop resistor
- Integrated pneumatic pressure pump up to 70 bar
- Pressure module: any scale from -0.95 to 70 bar
- Accuracy: 0.01 % of full scale
- Temperature compensation up to 50 °C
- High stability & accuracy
- Simulates & measures electrical signals
- 4.3 " TFT colour touch screen
- 8 hours battery life
- USB PC interface & Calibration software
- ISO17025 traceable calibration certificate
Stability: ± 0.02 °C
Resolution: 0.01 °C
Stabilization time: 7 minutes
Temperature unit: °C or °F selectable
Well diameter: 30 mm
Immersion length: 150 mm
Heating time: 10 minutes (Amb. to 120 °C)
Cooling time: 15 minutes (Amb. to (-) 10 °C)
25 minutes (120 °C to (-) 10 °C)
Communication port: RS232 port
Calibration software: MET-CAL as standard
Switch test: Provided
Power supply / Consumption: 110V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (W x D x H) / Weight: 160 x 365 x 450 mm / 9 kg
Type of case: Portable
Carrying case: Provided
ISO17025 calibration certificate: Provided
Température range : 50 °C ~ 650 °C
Accuracy: ± 0.3 °C
Stability: ± 0.02 °C
Resolution: 0.01 °C
Stabilization time: 7 minutes
Temperature unit: °C or °F selectable
Well diameter: 30 mm
Immersion length: 150 mm
Heating time: 10 minutes (Amb. to 120 °C)
Cooling time: 15 minutes (Amb. to (-) 10 °C)
25 minutes (120 °C to (-) 10 °C)
Communication port: RS232 port
Calibration software: MET-CAL as standard
Switch test: Provided
Power supply / Consumption: 110V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (W x D x H) / Weight: 160 x 365 x 450 mm / 9 kg
Type of case: Portable
Carrying case: Provided
ISO17025 calibration certificate: Provided
Accuracy: ± 0.3 °C
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 7 minutes
Temperature unit: Selectable °C or °F
Well diameter: 19 mm
Immersion length: 100 mm
Heating time: 14 minutes (ambient to 123 °C)
Cooling time: 34 minutes (123 °C to (-)40 °C)
Interchangeable insert: Well for Ø 1/4’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 100 V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 9 kg
Enclosure type: Portable
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 20 minutes
Temperature unit: Selectable °C or °F
Well diameter: 25 mm
Immersion length: 110 mm
Heating time: 20 minutes (ambient to 350 °C)
Cooling time: 33 minutes (350 °C to ambient temperature)
Interchangeable insert: Well for Ø 1/4’’ and 1/2’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 230 V AC, 50 Hz, 520 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 9 kg
Enclosure type: Portable
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 20 minutes
Temperature unit: Selectable °C or °F
Well diameter: 25 mm
Immersion length: 110 mm
Heating time: 20 minutes (ambient to 650 °C)
Cooling time: 42 minutes (650 °C to ambient temperature)
Interchangeable insert: Well for Ø 1/4’’ and 1/2’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 230 V AC, 50 Hz, 1230 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 11 kg
Enclosure type: Portable