- Pressure range: -0.9 to 60 bar
- Media: Air
- Resolution: 1 mbar
- Connections Standard instruments: 3/8 BSP F
- Connections Instruments test: 1/4 BSP F
- Nylon hose 1 meter with suitable connector
- Ergonomics & Ease of use
- Robust Design & Stable Performance
- No specific maintenance to be expected
Stability: ± 0.02 °C
Resolution: 0.01 °C
Stabilization time: 7 minutes
Temperature unit: °C or °F selectable
Well diameter: 30 mm
Immersion length: 150 mm
Heating time: 10 minutes (Amb. to 120 °C)
Cooling time: 15 minutes (Amb. to (-) 10 °C)
25 minutes (120 °C to (-) 10 °C)
Communication port: RS232 port
Calibration software: MET-CAL as standard
Switch test: Provided
Power supply / Consumption: 110V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (W x D x H) / Weight: 160 x 365 x 450 mm / 9 kg
Type of case: Portable
Carrying case: Provided
ISO17025 calibration certificate: Provided
Température range : 50 °C ~ 650 °C
Accuracy: ± 0.3 °C
Stability: ± 0.02 °C
Resolution: 0.01 °C
Stabilization time: 7 minutes
Temperature unit: °C or °F selectable
Well diameter: 30 mm
Immersion length: 150 mm
Heating time: 10 minutes (Amb. to 120 °C)
Cooling time: 15 minutes (Amb. to (-) 10 °C)
25 minutes (120 °C to (-) 10 °C)
Communication port: RS232 port
Calibration software: MET-CAL as standard
Switch test: Provided
Power supply / Consumption: 110V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (W x D x H) / Weight: 160 x 365 x 450 mm / 9 kg
Type of case: Portable
Carrying case: Provided
ISO17025 calibration certificate: Provided
Accuracy: ± 0.3 °C
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 7 minutes
Temperature unit: Selectable °C or °F
Well diameter: 19 mm
Immersion length: 100 mm
Heating time: 14 minutes (ambient to 123 °C)
Cooling time: 34 minutes (123 °C to (-)40 °C)
Interchangeable insert: Well for Ø 1/4’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 100 V to 240V AC, 50 to 60 Hz, 150 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 9 kg
Enclosure type: Portable
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 20 minutes
Temperature unit: Selectable °C or °F
Well diameter: 25 mm
Immersion length: 110 mm
Heating time: 20 minutes (ambient to 350 °C)
Cooling time: 33 minutes (350 °C to ambient temperature)
Interchangeable insert: Well for Ø 1/4’’ and 1/2’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 230 V AC, 50 Hz, 520 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 9 kg
Enclosure type: Portable
Stability: ± 0.05 °C
Resolution: 0.1 °C
Stabilization time: 20 minutes
Temperature unit: Selectable °C or °F
Well diameter: 25 mm
Immersion length: 110 mm
Heating time: 20 minutes (ambient to 650 °C)
Cooling time: 42 minutes (650 °C to ambient temperature)
Interchangeable insert: Well for Ø 1/4’’ and 1/2’’ probes as standard (other sizes available on request).
Switch test: Provided
Power supply / Consumption: 230 V AC, 50 Hz, 1230 W
Dimensions (L x W x H) / Weight: 160 x 365 x 350 mm / 11 kg
Enclosure type: Portable
- Accuracy: ± 1% of full scale
- 12 Units: kPa, Pa, PSI, Kg / cm², Bar, mBar, MPa, Pa ...
- Display: Large LCD screen with 4 DIGIT
- Keys: On / Off, Reset, backlight
- Connection: 1/4 ’’ G
- Dial size: 65 mm
- All stainless steel fitting and housing
- Max pressure recording & min.
- Battery alarm function
- Pressure diagram in%
- Ease of use, robust design and reliable performance
- Lithium battery 1 year autonomy in 24 hour operation
IN STOCK - Delivery within 48 hours max.
- 12 Units: kPa, Pa, PSI, Kg / cm², Bar, mBar, MPa, Pa ...
- Display: Large LCD screen with 4 DIGIT
- Keys: On / Off, Reset, backlight
- Connection: 1/4 ’’ G
- Dial size: 65 mm
- All stainless steel fitting and housing
- Max pressure recording & min.
- Battery alarm function
- Pressure diagram in%
- Ease of use, robust design and reliable performance
- Lithium battery 1 year autonomy in 24 hour operation
In process of supply - delivery within 10 days
- Accuracy: ± 0.2% of full scale
- Units: kPa, Pa, PSI, Kg / cm², Bar, MPa
- Display: Large LCD display with 5 DIGIT + Pressure bar graph
- Keys: On / Off, Reset, backlight
- Connection: 1/4 ’’ NPT
- Dial size: 80 mm
- Max & min pressure recording
- Battery alarm function
- Ease of use, robust design and reliable performance
- 2x Lithium battery
CALIBRATION CERTIFICATE >>
- Accuracy: ± 0.4% of full scale
- Units: kPa, Pa, PSI, Kg / cm², Bar, MPa
- Display: Large LCD display with 5 DIGIT + Pressure bar graph
- Keys: On / Off, Reset, backlight
- Connection: 1/4 ’’ NPT
- Dial size: 80 mm
- Max & min pressure recording
- Battery alarm function
- Ease of use, robust design and reliable performance
- 2x Lithium battery
CALIBRATION CERTIFICATE >>